Ultrastructural Analysis of the Surface of Endodontic Instruments after Immersion in Irrigating Solutions | Ultrastrukturna analiza površine endodontskih instrumenata nakon potapanja u rastvore za irigaciju |
Serbian Dental Journal, Original article [pdf] | Stomatološki Glasnik Srbije, Originalni rad [pdf] |
ID: 4.2015.4.1 Number: 4 Volume: 62 Month: 12 Year: 2015 UDC: 620.193.4:616.314-78 [tmx] [bow] |
Jelena Popović Institution: University of Niš, Medical Faculty, Clinic of Dentistry, Department of Restorative Dentistry and Endodontics, Niš, Serbia Mail: jelenadp@gmail.com | Jelena Popović Institucija: Univerzitet u Nišu, Medicinski fakultet, Klinika za stomatologiju, Odeljenje za bolesti zuba i endodonciju, Niš, Srbija E-pošta: jelenadp@gmail.com |
Goran Radenković Institution: University of Niš, Mechanical Faculty, Department of Production Engineering, Niš, Serbia | Goran Radenković Institucija: Univerzitet u Nišu, Mašinski fakultet, Katedra za proizvodno-informacione tehnologije i menadžment, Niš, Srbija |
Jovanka Gašić Institution: University of Niš, Medical Faculty, Clinic of Dentistry, Department of Restorative Dentistry and Endodontics, Niš, Serbia | Jovanka Gašić Institucija: Univerzitet u Nišu, Medicinski fakultet, Klinika za stomatologiju, Odeljenje za bolesti zuba i endodonciju, Niš, Srbija |
Aleksandar Mitić Institution: University of Niš, Medical Faculty, Clinic of Dentistry, Department of Restorative Dentistry and Endodontics, Niš, Serbia | Aleksandar Mitić Institucija: Univerzitet u Nišu, Medicinski fakultet, Klinika za stomatologiju, Odeljenje za bolesti zuba i endodonciju, Niš, Srbija |
Marija Nikolić Institution: University of Niš, Medical Faculty, Clinic of Dentistry, Department of Restorative Dentistry and Endodontics, Niš, Serbia | Marija Nikolić Institucija: Univerzitet u Nišu, Medicinski fakultet, Klinika za stomatologiju, Odeljenje za bolesti zuba i endodonciju, Niš, Srbija |
Radomir Barac Institution: University of Niš, Medical Faculty, Clinic of Dentistry, Department of Restorative Dentistry and Endodontics, Niš, Serbia | Radomir Barac Institucija: Univerzitet u Nišu, Medicinski fakultet, Klinika za stomatologiju, Odeljenje za bolesti zuba i endodonciju, Niš, Srbija |
Slavoljub Živković Institution: University of Belgrade, Faculty of Dental Medicine, Department of Restorative Dentistry and Endodontics, Belgrade, Serbia | Slavoljub Živković Institucija: Univerzitet u Beogradu, Stomatološki fakultet, Klinika za bolesti zuba i endodonciju, Beograd, Srbija |
Abstract Introduction
Separation (fracture) of endodontic instruments in the root canal during chemomechanical instrumentation is a complication that can compromise the final outcome of endodontic treatment. One of the most common factors that cause fatigue of endodontic instruments and consequent fracture is surface corrosion. The aim of this study was to investigate the ultrastructure of surface corrosion of endodontic instruments made of stainless steel and nickel-titanium after immersion in the most commonly used root canal irrigants.
Material and Methods
The study included 48 nickel-titanium and stainless steel endodontic hand files. All instruments were immersed in 5.25% sodium hypochlorite, 0.2% CHX and 17% EDTA. Surface corrosion was analyzed using a scanning electron microscope (SEM).
Results
Nickel-titanium instruments showed significantly higher susceptibility to corrosion after immersion in 5.25% sodium hypochlorite compared to stainless steel instruments (p<0,001). After immersion in 0.2% CHX corrosion damage was observed on both nickel-titanium and stainless steel instruments but the difference was not statistically significant (p=0.096). No corrosion was observed in both types of instruments after immersion in 17% EDTA.
Conclusion
The use of 5.25% NaOCl and 0.2% CHX as root canal irrigating solutions can cause serious corrosion changes on the surface of nickel-titanium and stainless steel endodontic instruments. | Apstrakt Uvod
Fraktura endodontskih instrumenata u kanalu korena tokom hemomehaničke preparacije je značajna komplikacija koja može ugroziti konačan ishod endodontskog lečenja. Jedan od važnih faktora koji izazivaju zamor materijala endodontskog instrumenta je površinska korozija. Cilj ovog rada je bio da se ispita ultrastruktura površine endodontskih instrumenata od nerđajućeg čelika i nikl-titanijuma nakon potapanja u najčešće upotrebljavane rastvore za irigaciju kanala korena zuba.
Materijal i metode rada
U istraživanju je korišćeno 48 instrumenata od nikl-titanijuma i nerđajućeg čelika. Testiranje osetljivosti na koroziju je izvršeno potapanjem u rastvore NaOCl od 5,25%, CHX od 0,2% i EDTA od 17%. Analiza korozivnih oštećenja je urađena pomoću skening elektronskog mikroskopa (SEM) na različitim uveličanjima.
Rezultati
Instrumenti od nikl-titanijuma su pokazali statistički značajno veću osetljivost na koroziju nakon potapanja u 5,25% NaOCl u poređenju sa instrumentima od nerđajućeg čelika (p<0,001). Nakon potapanja u 0,2% CHX uočeno je korozivno oštećenje instrumenata od nikl-titanijuma i nerđajućeg čelika, ali razlika nije bila statistički značajna. Korozija nakon potapanja u 17% EDTA nije uočena ni kod jedne vrste endodontskih instrumenata.
Zaključak
Primena 5,25% NaOCl i 0,2% CHX kao rastvora za irigaciju kanala korena može izazvati ozbiljne korozivne promene na površini instrumenata od nikl-titanijuma i nerđajućeg čelika. |
Keywords: corrosion; endodontic instruments; nickel-titanium; stainless steel; SEM | Ključne reči: korozija; endodontski instrumenti; nikl-titanijum; nerđajući čelik; SEM |
Pages: 157-165 | Strane: 157-165 |